Tolerance-Dependent Inspection Disagreement Between a Fixed CMM and a Portable Articulated-Arm CMM
Abstract
Fixed coordinate measuring machines (CMMs) and portable articulated-arm CMMs are often assigned to the same inspection task, but their nominal accuracy specifications do not show whether a change of instrument will preserve the disposition of a part. The question is not simply how far the two results differ, but whether that difference crosses the tolerance boundary. We examined this issue with recorded measurements of cylindrical, cubic, and spherical features under nominal 20 °C and 30 °C conditions. Repeated records and two roughness profiles without sufficient acquisition information were removed, leaving six dimensional and four form profiles. For each dimensional feature, the distances of the two system means from nominal define the exact tolerance interval in which the systems receive opposite direct labels. The fixed-CMM stream was approximately 11.2 μm higher than the articulated-arm stream at both conditions. All four form profiles fell on opposite sides of the recorded 10 μm upper limit. The dimensional disagreement intervals also overlapped strongly; their mean widths were 6.573 μm at 20 °C and 4.995 μm at 30 °C. The results clarify why an average difference between instruments is not, by itself, a measure of substitution risk. The proposed tolerance map identifies the feature-tolerance combinations for which instrument choice can change the recorded inspection label and, therefore, where a controlled equivalence study and a task-specific uncertainty budget are needed before substitution.
Read the original paper
This page indexes public paper metadata. The manuscript remains with its original publisher and authors.