Back to Research papers
Research paper index

Safety Analysis of Metasurface-Based Near-field Wireless Power Transfer System for Deep Implant

Maoyuan Li, Ali Khaleghi, Ilangko Balasingham

arXiv:2607.29187Published July 31, 20260 citations
  • eess.SP

Abstract

Wireless power transfer is a method for energizing future implantable medical electronics. In this study, a metasurface-based near-field magnetic wireless power transfer system for deep implants is presented, and electromagnetic safety parameters, including field distributions, specific absorption rate (SAR), and temperature variations, are evaluated. The power transfer is modeled for a receiver implant at a distance of 8.5 cm from the designed metasurface. Based on the results, a maximum localized SAR of 0.072 mW/kg is achieved when the efficiency is 1.62%. Moreover, continuous power transfer shows that the local tissue temperature rises by less than 1.1 degrees Celsius.

Read the original paper

This page indexes public paper metadata. The manuscript remains with its original publisher and authors.